Monitoring

Pump Features/Functions

Protection
Purge option (if enabled)

To reduce or eliminate slack (visible spaces at the syringe collar and plunger loading points) and ensure a faster start-up time (time to reach the programmed infusion rate), the user can purge the system (once only) up to a configured volume limit.

Accessing the Info menu

With no infusion running, press the INFO key.

Options for viewing are:
  • Battery Level / Exit / Rate Setting / Event Log / Change Set-up
  • Scroll menu using ↑↓ keys and press "YES" to view the option selected
Pre-Loading and Automatic Actuator Movement
  • Pre-Loading and automatic actuator movement commences when the pump is powered up with the barrel clamp arm down and no syringe is in place. NOTE: These sequences will not occur if the barrel clamp arm is raised when the pump is powered on.
  • During Pre-Loading, the display screens displays pump information and the actuator moves automatically.
  • These simultaneous sequences clear a programme if still in the pump memory.
  • At the end of Pre-Loading the actuator returns to the start position of the last infusion programmed.