When an ALERT is activated:
- The infusion continues
- 2/3 beeps are heard approximately every 3 to 4 minutes
- A screen message indicating the cause of the alert displays intermittently with the infusion running screen
Alert activates approximately 15 minutes prior to infusion end
When an ALARM is activated:
- The infusion stops
- A continuous audible alarm activates (this will continue until either the YES key is pressed to mute or the problem is rectified
- A screen message displays to indicate the cause of the alarm
- The infusion status indicator (LED) turns red