Programme Protection

Alarms

Programme protection applies specifically to the ml/hour rate

This screen prompt displays if the programme is interrupted by alarm activation (e.g. syringe displacement, occlusion) the pump is powered off for any reason and powered on with a syringe in place and on completion of purge:

Pressing "YES to Resume" retains the current programme (ml/hr rate)

Pressing "No for New Syringe" deletes the current programme enabling a new programme to be calculated using the syringe volume at that time.

The important feature to remember is that "Resume" protects the calculated infusion rate for the current syringe.

  • If you increase the syringe volume and resume the programme, the duration of delivery will increase
  • If you decrease the syringe volume and resume the programme, the duration of delivery will decrease