Purge option (if enabled)
To reduce or eliminate slack (visible spaces at the syringe collar and
plunger loading points) and ensure a faster start-up time (time to reach
the programmed infusion rate), the user can purge the system (once
only) up to a configured volume limit.
Accessing the Info menu
With no infusion running, press the
INFO key.
Options for viewing are:
- Battery Level / Exit / Rate Setting / Event Log / Change Set-up
- Scroll menu using ↑↓ keys and press "YES" to view the option selected
Pre-Loading and Automatic Actuator Movement
- Pre-Loading and automatic actuator movement commences when the pump is powered up with the barrel clamp arm down and no syringe is in place. NOTE: These sequences will not occur if the barrel clamp arm is raised when the pump is powered on.
- During Pre-Loading, the display screens displays pump information and the actuator moves automatically.
- These simultaneous sequences clear a programme if still in the pump memory.
- At the end of Pre-Loading the actuator returns to the start position of the last infusion programmed.